Please use this identifier to cite or link to this item:
http://acervodigital.unesp.br/handle/11449/7941
- Title:
- Development and validation of HPLC method for analysis of dexamethasone acetate in microemulsions
- Universidade Estadual Paulista (UNESP)
- Universidade Estadual do Centro-Oeste Departamento de Farmácia
- 1984-8250
- Rede Nanoblotech
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)
- Um método simples, rápido, preciso e sensível foi desenvolvido para a análise quantitativa de acetato de dexametasona em microemulsões usando cromatografia líquida de alta eficiência (CLAE). Os parâmetros cromatográficos foram: coluna cromatográfica Lichrospher 100 RP-18, (250 mm x 4 mm i.d., 5 μm partícula tamanho), com temperatura de coluna de 30 ± 2 ºC. A fase móvel foi composta de metanol: água (65:35; v/v) com fluxo isocrático de 1 mL.min-1 e volume de injeção de 20 μL. As determinações foram realizadas utilizando detector UV-Vis no comprimento de onda de 239 nm. A curva analítica mostrou-se linear (r² 0,999) em uma ampla faixa de concentração (2,0-30,0 μg.mL-1). A presença de componentes da microemulsão não interferiu nos resultados da análise. O método mostrou precisão adequada, com desvio padrão relativo menor que 3%. A exatidão foi analisada pela adição de padrões do fármaco e valores de recuperação dentro dos limites recomendáveis foram obtidos para todas as concentrações estudadas. O método por CLAE mostrou especificidade e seletividade com linearidade dentro da faixa de concentração utilizada e precisão e exatidão que tornam o método adequado para a análise de dexametasona em microemulsões. O procedimento analítico é fidedigno e oferece vantagens em termos de velocidade e custo de reativos.
- A simple, rapid, accurate and sensitive method was developed for quantitative analysis of dexamethasone acetate in microemulsions using high performance liquid chromatography (HPLC) with UV detection. The chromatography parameters were stainless steel Lichrospher 100 RP-18 column (250 mm x 4 mm i.d., 5 μm particle size), at 30 ± 2 ºC. The isocratic mobile phase was methanol:water (65:35; v/v) at a flow rate of at 1.0 mL.min-1. The determinations were performed using UV-Vis detector set at 239 nm. Samples were prepared with methanol and the volume injected was 20 μL. The analytical curve was linear (r² 0.9995) over a wide concentration range (2.0-30.0 μg.mL-1). The presence of components of the microemulsion did not interfere in the results of the analysis. The method showed adequate precision, with a relative standard deviation (RSD) smaller than 3%. The accuracy was analyzed by adding a standard drug and good recovery values were obtained for all drug concentrations used. The HPLC method developed in this study showed specificity and selectivity with linearity in the working range and good precision and accuracy, making it very suitable for quantification of dexamethasone in microemulsions. The analytical procedure is reliable and offers advantages in terms of speed and low cost of reagents.
- 1-Mar-2009
- Brazilian Journal of Pharmaceutical Sciences. Universidade de São Paulo, Faculdade de Ciências Farmacêuticas, v. 45, n. 1, p. 87-92, 2009.
- 87-92
- Universidade de São Paulo (USP), Faculdade de Ciências Farmacêuticas
- Acetato de dexametasona
- Microemulsões
- Cromatografia líquida de alta eficiência
- Método de análise
- Dexamethasone acetate
- Microemulsions
- High performance liquid chromatography
- Analytical method
- http://dx.doi.org/10.1590/S1984-82502009000100010
- http://hdl.handle.net/11449/7941
- Acesso aberto
- outro
- http://repositorio.unesp.br/handle/11449/7941
There are no files associated with this item.