Design of dental implants, influence on the osteogenesis and fixation
Author(s): 
Institution: 
  • Universidade Estadual Paulista (UNESP)
  • Natl Inst Technol
  • Aerosp Tech Ctr CTA
ISSN: 
0957-4530
Abstract: 
The fixation and the bone ingrowth at the interface of porous cylindrical implants (total porosity of 37% and average pores diameter of 480 mu m) were compared in vivo to rough cylindrical implants (R-a = 5.3 mu m), both of commercially pure titanium, made by powder metallurgy. The implants were inserted into the tibias of 20 rabbits and the animals were sacrificed 4 and 8 weeks after surgery. The percentage of bone-implant contact observed in porous implant was significantly larger than in the rough ones for all of sacrifice periods, respectively, 57% vs. 46% after 4 weeks, and 59% vs. 50% after 8 weeks. The mechanical tests showed a significant increase in the shear strength of the porous implants for the two analyzed periods, 4 and 8 weeks (14 and 20 MPa), when compared with rough ones (4 and 13 MPa). These results suggest that porous implants improve the contact at the implant-bone interface and increase the fixation to the bone, improving the osseointegration. Thus, the porous implant might be an alternative to dental implant in less favorable conditions, and appear to be better fixed to bone, offering promising alternatives.
Issue Date: 
1-Aug-2008
Citation: 
Journal of Materials Science-materials In Medicine. Dordrecht: Springer, v. 19, n. 8, p. 2851-2857, 2008.
Time Duration: 
2851-2857
Publisher: 
Springer
Source: 
http://dx.doi.org/10.1007/s10856-008-3421-6
URI: 
Access Rights: 
Acesso restrito
Type: 
outro
Source:
http://repositorio.unesp.br/handle/11449/22798
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.