Please use this identifier to cite or link to this item:
http://acervodigital.unesp.br/handle/11449/128409
- Title:
- Influence of calibration protocols for a pressure-sensing walkway on kinetic and temporospatial parameters
- Universidade Estadual Paulista (UNESP)
- Univ Fed Rural Amazônia (UFRA)
- 0932-0814
- Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)
- CAPES: 21/2009
- Objectives: To evaluate the influence on the kinetic and temporospatial parameters of calibration protocols with point and step techniques for a pressure-sensing walkway.Methods: Nine Labrador dogs were used. Two protocols of point calibration technique (Cl and C2) and eight protocols of step calibration technique (C3 to C10) were performed. In Cl, weight was added to a stool to match the body mass of each dog. In C2, weight was added to the stool to match a 46.1 kg person. The other eight calibration protocols represented combinations of the following factors: 46.1 kg and 96.1 kg persons, barefoot or wearing sneakers, and stepping onto the platform with one or two feet.Results: The calibration protocols did not affect the temporospatial variables or percentages of body weight (%BW) distribution. Significant differences were found in both PVI (peak vertical force) and VI (vertical impulse) between barefoot versus wearing sneakers, 46.1 kg versus 96.1 kg person, and stepping onto the platform with one foot versus two feet. When comparing Cl with other protocols, significant differences were observed in PVF and VI for both forelimbs and hindlimbs. When comparing C2 with other protocols, significant differences were observed in PVF and VI for both forelimbs and hindlimbs in all protocols.Clinical significance: The PVF and VI were influenced by the calibration protocol used, but the %BW distribution and temporospatial parameters were not. Using the same calibration protocol for all dogs within the same group eliminated the variability of the kinetic data caused by the calibration.
- 1-Jan-2015
- Veterinary And Comparative Orthopaedics And Traumatology. Stuttgart: Schattauer Gmbh-verlag Medizin Naturwissenschaften, v. 28, n. 1, p. 25-29, 2015.
- 25-29
- Schattauer Gmbh-verlag Medizin Naturwissenschaften
- Pressure platform
- Tekscan
- Calibration
- http://vcot.schattauer.de/en/contents/archivestandard/issue/2169/manuscript/23786.html
- Acesso aberto
- outro
- http://repositorio.unesp.br/handle/11449/128409
There are no files associated with this item.
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.