You are in the accessibility menu

Please use this identifier to cite or link to this item: http://acervodigital.unesp.br/handle/11449/129086
Title: 
lmmittance electroanalysis in diagnostics
Author(s): 
Institution: 
  • University of Oxford
  • Universidade Estadual Paulista (UNESP)
ISSN: 
0003-2700
Sponsorship: 
  • Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
  • Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
Abstract: 
Impedance derived electroanalytical assays are inherently spectroscopic (frequency resolved) and potentially exceedingly sensitive indicators of interfacial change (such as target binding at an appropriate receptor). We introduce here the use of a portfolio of mathematically derived immittance functions and related components, capable, from the same raw data sets, of enabling increased assay sensitivity and markedly shorter assay times in comparison to traditional impedance analyses. The methodology, applied herein to faradaic (redox probe amplified) and non-faradaic assays, requires no equivalent circuit analysis or prior assumption of response. Its focus is to optimize analytical potency and to enable the user to select and apply the most frequency-optimized reporter of interfacial change and to, thereafter, run rapid (optimized) analyses at single frequencies.
Issue Date: 
20-Jan-2015
Citation: 
Analytical Chemistry. Washington: Amer Chemical Soc, v. 87, n. 2, p. 944-950, 2015.
Time Duration: 
944-950
Publisher: 
Amer Chemical Soc
Source: 
http://pubs.acs.org/doi/abs/10.1021/ac503156a
URI: 
Access Rights: 
Acesso restrito
Type: 
outro
Source:
http://repositorio.unesp.br/handle/11449/129086
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.