You are in the accessibility menu

Please use this identifier to cite or link to this item: http://acervodigital.unesp.br/handle/11449/18997
Title: 
Differential pulse voltammetric determination of nitrite with gold ultramicroelectrode
Author(s): 
Institution: 
  • Universidade Estadual Paulista (UNESP)
  • Universidade de São Paulo (USP)
ISSN: 
1040-0397
Abstract: 
The application of disk shaped gold ultramicroelectrode for nitrite determination with and without addition of supporting electrolyte was studied using the differential pulse voltammetric method. The well-defined peak for nitrite oxidation near 0.8V (vs. Ag/AgCl reference electrode) was used to obtain analytical plots in the concentration range from 0.1 to 0.6 mmol L-1 and from 10.0 to 50.0 mu mol.L-1. The calculated detection limit was 0.65 mu mol.L-1 in purified water, in the absence of supporting electrolyte, with relative standard deviation of 1.36% (n=6) for analyzing 10.0 mu mol L-1 nitrite solutions, and accuracy of 100.9 %, based on recovery studies. The application of this analytical method to mineral and river water samples of natural pH also showed improved sensitivity when compared with the linear sweep voltammetric method previously reported.
Issue Date: 
1-Dec-1998
Citation: 
Electroanalysis. Berlin: Wiley-v C H Verlag Gmbh, v. 10, n. 17, p. 1200-1203, 1998.
Time Duration: 
1200-1203
Publisher: 
Wiley-Blackwell
Keywords: 
  • nitrite
  • ultramicroelectrodes
  • natural water
  • differential pulse voltammetry
Source: 
http://dx.doi.org/10.1002/(SICI)1521-4109(199811)10:17<1200