Internal standardization and least-squares background correction in high-resolution continuum source flame atomic absorption spectrometry to eliminate interferences on determination of Pb in phosphoric acid
Author(s): 
Institution: 
  • Universidade Estadual Paulista (UNESP)
  • Universidade Federal de São Carlos (UFSCar)
ISSN: 
0584-8547
Sponsorship: 
  • Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
  • Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
Abstract: 
In this work the least-squares background correction (LSBC) and internal standardization (IS) techniques were combined to eliminate spectral and transport interferences in the determination of Pb in phosphoric acid by high-resolution continuum source atomic absorption spectrometry (HR-CS AAS). Blanks, samples and reference solutions [0.10-1.00 mg L-1 Ph in 1% (v/v) HNO3] were spiked with 4.00 mg L-1 Co used as internal standard. For absorbance measurements at the wavelength integrated absorbance equivalent to 9 pixels, correlations between the ratio of absorbance of Ph to absorbance of Co and the analyte concentration were close to 0.9992. Relative standard deviations of measurements varied from 0.6 to 4% and 1 to 7% (n = 12) without and with IS/LSBC techniques, respectively. Recoveries for Ph spikes were in the 96-104% and 76-180% range with and without IS/LSBC, respectively. The limit of detection improved with IS/LSBC techniques. Accuracy of the proposed method was checked for the determinations of Ph in commercial phosphoric acid samples and results obtained with IS were better than those without IS. (C) 2008 Elsevier B.V. All rights reserved.
Issue Date: 
1-Sep-2008
Citation: 
Spectrochimica Acta Part B-atomic Spectroscopy. Oxford: Pergamon-Elsevier B.V. Ltd, v. 63, n. 9, p. 992-995, 2008.
Time Duration: 
992-995
Publisher: 
Pergamon-Elsevier B.V. Ltd
Keywords: 
  • Lead
  • Phosphoric acid
  • Least Least-squares background correction
  • Internal standardization
  • High-resolution continuum source atomic
  • absorption spectrometry
Source: 
http://dx.doi.org/10.1016/j.sab.2008.05.003
URI: 
Access Rights: 
Acesso restrito
Type: 
outro
Source:
http://repositorio.unesp.br/handle/11449/25783
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.