You are in the accessibility menu

Please use this identifier to cite or link to this item: http://acervodigital.unesp.br/handle/11449/40053
Title: 
Heat exchangers for cardioplegia systems: In vitro study of four different concepts
Author(s): 
Institution: 
  • Universidade Estadual de Campinas (UNICAMP)
  • Universidade Estadual Paulista (UNESP)
ISSN: 
0160-564X
Abstract: 
The aim of this work is the evaluation of four different heat exchangers used for myocardium during cardioplegic system in cardiac surgeries. Four types of shell and tube heat exchangers made of different exchange elements were constructed, as follows: stainless steel tubes, aluminium tubes, polypropylene hollow fiber, and bellows type. The evaluation was performed by in vitro tests of parameters such as heat transfer, pressure drop, and hemolysis tendency. The result has shown that all four systems tested were able to achieve the heat performance, and to offer low resistance to flow, and safety, as well as have low tendency to hemolysis. However, we can emphasize that the bellows type heat exchanger has a significant difference with regard to the other three types.
Issue Date: 
1-May-2003
Citation: 
Artificial Organs. Malden: Blackwell Publishing Inc., v. 27, n. 5, p. 465-468, 2003.
Time Duration: 
465-468
Publisher: 
Blackwell Publishing
Keywords: 
  • blood cardioplegia
  • heat exchangers
  • myocardial preservation
Source: 
http://dx.doi.org/10.1046/j.1525-1594.2003.07240.x
URI: 
Access Rights: 
Acesso restrito
Type: 
outro
Source:
http://repositorio.unesp.br/handle/11449/40053
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.