Please use this identifier to cite or link to this item:
http://acervodigital.unesp.br/handle/11449/64933
- Title:
- Determination of Oxalate in Grass by FIA with a Spectrophotometric Detection System Using a Two Enzyme Reactor
- Universidade Estadual de Campinas (UNICAMP)
- Universidade Estadual Paulista (UNESP)
- 0910-6340
- A new methodology for soluble oxalic acid determination in grass samples was developed using a two enzyme reactor in an FIA system. The reactor consisted of 3 U of oxalate oxidase and 100 U of peroxidase immobilized on Sorghum vulgare seeds activated with glutaraldehyde. The carbon dioxide was monitored spectrophotometrically, after reacting with an acid-base indicator (Bromocresol Purple) after it permeated through a PTFE membrane. A linear response range was observed between 0.25 and 1.00mmol l-1 of oxalic acid; the data was fit by the equation A=-0.8(±1.5)+ 57.2(±2.5)[oxalate], with a correlation coefficient of 0.9971 and a relative standard deviation of 2% for n=5. The variance for a 0.25 mmol l-1 oxalic acid standard solution was lower than 4% for 11 measurements. The FIA system allows analysis of 20 samples per hour without prior treatment. The proposed method showed a good correlation with that of the Sigma Kit.
- 1-Dec-1996
- Analytical Sciences, v. 12, n. 3, p. 443-447, 1996.
- 443-447
- Grass
- Oxalate determination
- Oxalate oxidase
- Sorghum vulgare seeds
- http://dx.doi.org/10.2116/analsci.12.443
- Acesso aberto
- outro
- http://repositorio.unesp.br/handle/11449/64933
There are no files associated with this item.