You are in the accessibility menu

Please use this identifier to cite or link to this item: http://acervodigital.unesp.br/handle/11449/66017
Title: 
Estudio de la pasivacion de la aleacion Co-Cr-Mo en suero fisiologico
Other Titles: 
Study of passivation of Co-Cr-Mo alloy in physiologic serum
Author(s): 
Institution: 
Universidade Estadual Paulista (UNESP)
ISSN: 
0716-8756
Abstract: 
Results are presented on the mechanism of passivation of Co-Cr-Mo biological implant alloys in physiological serum using open circuit potentiometry, potentiodynamic curves, and electrochemical impedance spectroscopy. The potential dependence of impedance data and the analysis of the parameters obtained indicate a progressive diminution of the initial layer thickness and the simultaneous formation of a second higher resistive layer. In more severe conditions than the existent in human body, the metallographic examination of the alloy surface shows localized corrosion in interdendritric regions. Elemental analysis of the surface reveals the presence of higher chromium content in these regions. The presence of chlorine was not detected, which suggested that during preferential attack, soluble species are also formed.
Issue Date: 
1-Dec-1999
Citation: 
Informacion Tecnologica, v. 10, n. 5, p. 61-65, 1999.
Time Duration: 
61-65
Keywords: 
  • Electrochemistry
  • Implants (surgical)
  • Passivation
  • Spectroscopic analysis
  • Open circuit potentiometry
  • Potentiodynamic curves
  • Cobalt alloys
URI: 
Access Rights: 
Acesso restrito
Type: 
outro
Source:
http://repositorio.unesp.br/handle/11449/66017
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.