- Title:
- Validation of microbiological assay for determination of cefuroxime in injectable preparations
- Universidade Estadual Paulista (UNESP)
- 0326-2383
- The validation of a microbiological assay, applying agar diffusion method for determination of the active of cefuroxime in power for injection, is described. Using a strain of Micrococcus luteus ATCC 9341 as the test organism, cefuroxime was measured in concentrations ranging from 30.0 to 120.0 μg/mL. The method validation showed that it is linear (r = 0.9999), precise (relative standard deviation = 0.37%) and accurate (it measured the added quantities). Microbiological assay is satisfactory for quantitation of cefuroxime in powder for injection and the validity of the proposed bioassay, which is a simple and a useful alternative methodology for cefuroxime determination in routine quality control.
- 22-Aug-2012
- Latin American Journal of Pharmacy, v. 31, n. 5, p. 746-750, 2012.
- 746-750
- Cefuroxime
- Cephalosporins
- Microbiological assay
- Quality control
- Validation
- cefuroxime
- agar diffusion
- bacterial growth
- bacterial strain
- bacterium culture
- calibration
- correlation coefficient
- drug determination
- drug purity
- drug structure
- limit of detection
- limit of quantitation
- microbiology
- Micrococcus luteus
- quality control
- validation study
- http://www.latamjpharm.org/resumenes/31/5/LAJOP_31_5_2_2.pdf