You are in the accessibility menu

Please use this identifier to cite or link to this item: http://acervodigital.unesp.br/handle/11449/75571
Title: 
Programação, conexão e avaliação de psicrômetro aspirado de termopar de baixo custo em micrologger para estudos agrometeorológicos
Other Titles: 
Programming, connection and avaliation of a low cost aspirated thermocouple psychrometer in micrologger for agrometeorological studies
Author(s): 
da Cunha, Antonio Ribeiro
Institution: 
Universidade Estadual Paulista (UNESP)
ISSN: 
  • 0102-7786
  • 1982-4351
Abstract: 
As the psychrometric method is considered standard for the measurement of relative humidity, many studies have used aspirated thermocouple psychrometers connected to microloggers but, however, they do not detail how the aspiration of the air stream is done. However, one of the major difficulties of aspirated thermocouple psychrometers is to program and to connect it in micrologger, because specific programming and connection are requires, and still some skills in getting accurate measurements. This work aimed to provide the programming and the connection of the aspirated thermocouple psychrometer type T for the CR10X microloggers and CR23X of Campbell Scientific, evaluating the quality of measurements of temperature and air relative humidity in relation the a sensor Vaisala HMP50. The non-continuous measurements were made in the period 2006/09/04 to 2007/07/11 at Jaboticabal, SP. The air temperature measurements were similar between the Vaisala sensors and aspirated thermocouple psychrometer, but the relative humidity measurements were significantly different. The measurements obtained by the aspirated thermocouple psychrometer connected to CR10X micrologger were the most accurate. Using the programming and connection of a micro-fan to suck in the dry and moist bulbs of aspirated thermocouple psychrometer in microloggers models CR10X and CR23X, it is possible to obtain measurements of air temperature with good precision and accuracy, and measurements of air relative humidity with good precision, but accuracy is not always attainable due to the difficulties inherent to the physical processes that occur in the wet bulb, the skill in the use and maintenance of this equipment and the reservoir protection against radiative effects.
Issue Date: 
1-Jun-2013
Citation: 
Revista Brasileira de Meteorologia, v. 28, n. 2, p. 221-232, 2013.
Time Duration: 
221-232
Keywords: 
  • Air temperature and relative humidity
  • Automatization
  • Micrologger
  • Thermocouple
Source: 
http://dx.doi.org/10.1590/S0102-77862013000200011
URI: 
Access Rights: 
Acesso aberto
Type: 
outro
Source:
http://repositorio.unesp.br/handle/11449/75571
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 