dc.identifier.citationPONTES, Wesley. Implementação de um sistema eletrônico para medir taxa de fluência de energia de radiação na faixa de radioterapia com sensor piroelétrico. 2011.. 91 f. Tese (doutorado) - Universidade Estadual Paulista, Faculdade de Engenharia de Ilha Solteira, 2011..-
dc.identifier.urihttp://hdl.handle.net/11449/103394-
dc.identifier.urihttp://acervodigital.unesp.br/handle/11449/103394-
dc.description.abstractSensores piroelétricos podem ser utilizados para medir a taxa de fluência de energia (intensidade) da radiação X na faixa de diagnóstico médico e de ortovoltagem. Neste trabalho, foi implementado um sistema eletrônico que pode medir taxa de fluência de energia de radiação gama de 1,25 MeV e fótons X de 6 MV e 15 MV. O sistema é constituído por um sensor piroelétrico, um conversor corrente-tensão de alta sensibilidade e um instrumento para registrar a forma de onda e medir o pulso do sinal produzido pelo sensor. Este instrumento pode ser um osciloscópio, um circuito microcontrolado associado a um display de cristal líquido, ou um módulo de aquisição de dados e um laptop. O sensor piroelétrico e o conversor corrente-tensão foram alojados em uma câmara piroelétrica blindada. Seis tipos de sensores piroelétricos foram avaliados, sendo três cerâmicas e três compósitos. Os resultados obtidos com o sistema eletrônico, utilizando os diferentes tipos de sensores, possibilitaram caracterizá-lo em termos de precisão, resolução e sensibilidade. O sistema apresentou resposta linear com a taxa de fluência de energia da radiação, precisão melhor que 3 % e resolução de 5,56 nW/m2. As constantes piezoelétricas d33 dos sensores utilizados não sofreram alteração após os mesmos terem sido irradiados com 45 Gy de radiação na faixa de radioterapia. Estas características, associadas à importância de se medir a taxa de fluência de energia de radiação ionizante, podem tornar o instrumento piroelétrico muito útil no monitoramento de radiação produzida por equipamentos de radioterapiapt
dc.description.abstractPyroelectric sensors have been be used to measure the energy fluence rate (intensity) of x-radiation in the range of medical diagnosis and orthovoltage. In this work, we implemented an electronic system that can measure the energy fluence rate of 1.25 MeV gamma radiation, 6 MV and 15 MV X-photons. The system consists of a pyroelectric sensor, a high sensitivity current-to-voltage converter and an instrument for measuring and recording the waveform of pulses produced by the pyroelectric sensor. This instrument may be an oscilloscope, a microcontrolled circuit associated with a liquid crystal display, or a data acquisition module connected to a laptop. The pyroelectric sensor and the current-tovoltage converter were placed in an shielded pyroelectric camera. Six types of pyroelectric sensors were evaluated, three ceramics and three composites. The results obtained with the electronic system, using different types of sensors, allowed us to characterize it in terms of precision, resolution and sensitivity. The system has linear response to the energy fluence rate of radiation, precision better than 3 % and resolution of 5,6 nW/m2. The piezoelectric constant d33 of the sensors used have not changed after these have been irradiated with 45 Gy of radiotherapy radiation. These characteristics, associated with the importance of measuring the ionizing radiation energy fluence rate may make the pyroelectric instrument very useful in monitoring radiation produced by radiotherapy equipmentsen
dc.description.sponsorshipCoordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)-
dc.format.extent91 f. : il.-
dc.language.isopor-
dc.publisherUniversidade Estadual Paulista (UNESP)-
dc.sourceAleph-
dc.subjectRadioterapiapt
dc.subjectRaios Xpt
dc.subjectRaios gamapt
dc.subjectRadiação - Intensidadept
dc.subjectFotonspt
dc.subjectSensor piroelétricopt
dc.subjectTaxa de fluência de energiapt
dc.subjectFótons de alta energiapt
dc.subjectPyroeletric sensoren
dc.subjectRadiotherapyen
dc.subjectX-rayen
dc.subjectGamma rayen
dc.subjectEnergy fluence rateen
dc.subjectRadiation intensityen
dc.subjectHigh energy photonsen
dc.titleImplementação de um sistema eletrônico para medir taxa de fluência de energia de radiação na faixa de radioterapia com sensor piroelétricopt
dc.typeoutro-
dc.contributor.institutionUniversidade Estadual Paulista (UNESP)-
dc.rights.accessRightsAcesso aberto-
dc.identifier.filepontes_w_dr_ilha.pdf-
dc.identifier.aleph000641061-
dc.identifier.capes33004099080P0-
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.