You are in the accessibility menu

Please use this identifier to cite or link to this item: http://acervodigital.unesp.br/handle/11449/36269
Full metadata record
DC FieldValueLanguage
dc.contributor.authorDe Lima, J. A.-
dc.contributor.authorCordeiro, A. S.-
dc.date.accessioned2014-05-20T15:25:57Z-
dc.date.accessioned2016-10-25T18:00:34Z-
dc.date.available2014-05-20T15:25:57Z-
dc.date.available2016-10-25T18:00:34Z-
dc.date.issued2002-05-01-
dc.identifierhttp://dx.doi.org/10.1109/10.995689-
dc.identifier.citationIEEE Transactions on Biomedical Engineering. New York: IEEE-Inst Electrical Electronics Engineers Inc., v. 49, n. 5, p. 497-500, 2002.-
dc.identifier.issn0018-9294-
dc.identifier.urihttp://hdl.handle.net/11449/36269-
dc.identifier.urihttp://acervodigital.unesp.br/handle/11449/36269-
dc.description.abstractA constant-current stimulator for high-impedance loads using only low-cost standard high-voltage components Is presented. A voltage-regulator powers an oscillator built across the primary of a step-up transformer whose secondary supplies, after rectification, the high voltage to a switched current-mirror in the driving stage. Adjusting the regulated voltage controls the pulsed-current intensity. A prototype produces stimulus of amplitude and pulsewidth within 0 less than or equal to I-skin less than or equal to 20 mA and 50 mus less than or equal to T-pulse less than or equal to 1 ms, respectively. Pulse-repetition spans from 1 Hz to 10 Hz. Worst case ripple is 3.7% at I-skin = 1 mA. Overall consumption is 5.6 W at I-skin = 20 mA.en
dc.format.extent497-500-
dc.language.isoeng-
dc.publisherInstitute of Electrical and Electronics Engineers (IEEE)-
dc.sourceWeb of Science-
dc.subjectcurrent sourcept
dc.subjectcurrent stimulatorpt
dc.subjectelectrocutaneous stimulationpt
dc.subjecthigh voltagept
dc.titleA low-cost neurostimulator with accurate pulsed-current controlen
dc.typeoutro-
dc.contributor.institutionUniversidade Estadual Paulista (UNESP)-
dc.description.affiliationUniv Estadual Paulista, Dept Elect Engn, VLSI Design & Instrumentat Grp, BR-12516410 Guaratingueta, SP, Brazil-
dc.description.affiliationUnespUniv Estadual Paulista, Dept Elect Engn, VLSI Design & Instrumentat Grp, BR-12516410 Guaratingueta, SP, Brazil-
dc.identifier.doi10.1109/10.995689-
dc.identifier.wosWOS:000174969600012-
dc.rights.accessRightsAcesso restrito-
dc.relation.ispartofIEEE Transactions on Biomedical Engineering-
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.