You are in the accessibility menu

Please use this identifier to cite or link to this item: http://acervodigital.unesp.br/handle/11449/69345
Full metadata record
DC FieldValueLanguage
dc.contributor.authorSilva, A. A.-
dc.contributor.authorFormariz, T. P.-
dc.contributor.authorScarpa, M. V.-
dc.contributor.authorOliveira, Anselmo Gomes de-
dc.date.accessioned2014-05-27T11:22:05Z-
dc.date.accessioned2016-10-25T18:23:12Z-
dc.date.available2014-05-27T11:22:05Z-
dc.date.available2016-10-25T18:23:12Z-
dc.date.issued2006-12-01-
dc.identifierhttp://serv-bib.fcfar.unesp.br/seer/index.php/Cien_Farm/article/viewArticle/371-
dc.identifier.citationRevista de Ciencias Farmaceuticas Basica e Aplicada, v. 27, n. 2, p. 119-126, 2006.-
dc.identifier.issn1808-4532-
dc.identifier.urihttp://hdl.handle.net/11449/69345-
dc.identifier.urihttp://acervodigital.unesp.br/handle/11449/69345-
dc.description.abstractA simple, rapid, selective and specific high performance liquid chromatographic (HPLC) method for quantitative analysis of the triamcinolone in polylactide-co-glycolide acid (PLGA) microparticles was developed. The chromatographic parameters were reversed-phase C18 column, 250mm x 4.6mm, with particle size 5 μm. The column oven was thermostated at 35°C ± 2°C. The mobile phase was methanol/water 45:55 (v/v) and elution was isocratic at a flow-rate of 1 mL.mL-1. The determinations were performed using a UV-Vis detector at 239 nm. The injected sample volume was 10 μL. The standard curve was linear (r2 > 0.999) in the concentration range 100-2500 ng.mL-1. The method showed adequate precision, with a relative standard deviation (RSD) was smaller than 3%. The accuracy was analyzed by adding a standard drug and good recovery values were obtained for all drug concentrations used. The method showed specificity and selectivity with linearity in the working range and good precision and accuracy, making it very suitable for quantitation of triamcinolone in PLGA microparticles.en
dc.format.extent119-126-
dc.language.isoeng-
dc.sourceScopus-
dc.subjectAnalytical method validation-
dc.subjectHPLC analytical method-
dc.subjectPLGA microparticles-
dc.subjectTriamcinolone-
dc.subjectglycolic acid derivative-
dc.subjectmethanol-
dc.subjecttriamcinolone-
dc.subjectwater-
dc.subjectbiodegradability-
dc.subjectconcentration (parameters)-
dc.subjectdiagnostic accuracy-
dc.subjectdiagnostic value-
dc.subjectelution-
dc.subjectflow rate-
dc.subjecthigh performance liquid chromatography-
dc.subjectinjection-
dc.subjectparticle size-
dc.subjectquantitative analysis-
dc.subjectsample-
dc.subjectstandard-
dc.subjectthermostability-
dc.subjectvalidation study-
dc.titleDevelopment and validation of HPLC method for quantitative analysis of triamcinolone in biodegradable microparticlesen
dc.typeoutro-
dc.contributor.institutionUniversidade Estadual Paulista (UNESP)-
dc.description.affiliationDepartamento de Fármacos e Medicamentos Faculdade de Ciências Farmacêuticas Universidade Estadual Paulista, UNESP, Araraquara, SP-
dc.description.affiliationDepartamento de Fármacos e Medicamentos Faculdade de Ciências Farmacêuticas Universidade Estadual Paulista, UNESP, Rodovia Araraquara-Jau, km. 01, CEP: 14801-902 - Araraquara - SP-
dc.description.affiliationUnespDepartamento de Fármacos e Medicamentos Faculdade de Ciências Farmacêuticas Universidade Estadual Paulista, UNESP, Araraquara, SP-
dc.description.affiliationUnespDepartamento de Fármacos e Medicamentos Faculdade de Ciências Farmacêuticas Universidade Estadual Paulista, UNESP, Rodovia Araraquara-Jau, km. 01, CEP: 14801-902 - Araraquara - SP-
dc.rights.accessRightsAcesso aberto-
dc.identifier.file2-s2.0-35348962072.pdf-
dc.relation.ispartofRevista de Ciências Farmacêuticas Básica e Aplicada-
dc.identifier.scopus2-s2.0-35348962072-
Appears in Collections:Artigos, TCCs, Teses e Dissertações da Unesp

There are no files associated with this item.
 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.